首页产品解决方案 / 营养泵

 

 

 

产品信息表

Section Content
Product Name Enteral Feeding Pump EPC Series
Variants Single Channel / Double Channel
Feature 1 8 hours battery long back up
Feature 2 Portable & Unique design with only 0.9KG
Feature 3 4.3 inch Big Touch screen, smart and light, wide range applications, interface friendly, operation easily to increase efficiency
Feature 4 Continuous Mode, intermittent Mode, Program Control Mode
Feature 5 DPS, monitoring the pressure changes in real time for automatic water flush to prevent occlusion of feeding set
Feature 6 More than 14 groups of nutrition solution already programmed into the pump, and can be chosen according to clinical use
Feature 7 Auto flushing, Auto water replenishment, Night mode, LCD brightness (9 level), etc. all these functions can be modified according to clinical use
- Continuous Mode / Intermittent Mode / Program Control Mode

Feeding Mode Parameter Table

Parameters Model EPC110 EPC112 EPC120 EPC122
Used for Single Bag Enteral Feeding Single Bag Enteral Feeding Double Bag Enteral Feeding Double Bag Enteral Feeding
Warmer NO 40±5°C NO 40±5°C
Set Compatibility Open System Open System Close System Close System
Feeding Accuracy ±5% after calibration ±5% after calibration ±5% ±5%
Flush Rate / / 1 - 800ml/h, step by 1ml/h 1 - 800ml/h, step by 1ml/h
Flush Volume / / 10 - 50ml, step by 1ml 10 - 50ml, step by 1ml
Water Replenishment Interval / / 00:01 - 24:00 (Hour:Minute) 00:01 - 24:00 (Hour:Minute)
Occlusion Level / / 1 - 11 level (20 - 120KPa ±15KPa) 1 - 11 level (20 - 120KPa ±15KPa)
Auto - Flush Pressure Level / / 1 - 9 level (20 - 100KPa ±15KPa) 1 - 9 level (20 - 100KPa ±15KPa)
Feeding Mode Continuous Feeding, Intermittant Feeding, Program Control Feeding
Range of Flow Rate 1 - 800ml/h, step by 1ml/h
Range of VTBI 0 - 9999ml, step by 1ml
Total Volume Infused 0 - 9999ml, <100ml, step by 0.1ml; ≥100ml, step by 1ml
Feeding Time Preset 00:01 - 999:59 (Hour: Minute)
Intermittant Feeding Interval 00:01 - 24:00 (Hour:Minute)
Intermittant Feeding Times 2 - 99 can be adjusted, step by 1
Prime/Bolus Rate 1 - 800ml/h, step by 1ml/h, defaulted rate 800ml/h
Reverse Rate 1 - 800ml/h, step by 1ml/h, default rate 200ml/h
Noise ≤ 50dB (A) , working on desk and 1m distance
Alarms and Prompts Occlusion, Completion, Near End, Feeding Mal - function, Pause overtime, Low Battery, Battery Exhausted, Motor Mal - function, AC Power disconnection, AC Power connection Indication, Battery Charging Indication, Battery Charging Full Indication