| Parameter | Details |
|---|---|
| I.V.Set Compatibility | Open System, match all brands of qualified PVC, TPE I.V.sets with external diameter of 3.8mm - 4.2mm |
| Range of Flow Rate | 1 - 1200 ml/h |
| Infusion Mode | Rate, Rate - Time, Rate - Volume, Time - Volume, Drop - Volume, Drop - Time |
| Purge/Bolus Rate | 1 - 1200 ml/h, default 800 ml/h, step by 1 ml/h |
| Single Bolus Volume | 0.1 - 10 ml, Adjustable, default 3 ml |
| Time Preset | 00:01~99:59 (Hour: Min) |
| Volume Range | 1 - 9999 ml |
| Drop Range | 1 - 400 d/min, step by 1 drop |
| Accuracy | ± 5% |
| Total Volume Infused | 0 - 9999 ml |
| Occlusion Pressure - High | 40 KPa ± 20 KPa |
| Occlusion Pressure - Middle | 60 KPa ± 20 KPa |
| Occlusion Pressure - Low | 100 KPa ± 20 KPa |
| Air in line Detection | Ultrasonic wave |
| Visible & Audible Alarms | Air in line, Down Stream Occlusion, Door Open, VTBI Completion, Near Completion, Low Battery, Battery Exhausted, Low Temperature, Motor Malfunction, A.C Disconnection, A.C connection, I.V set dislocation, MPU Error, Battery Charging, Battery Charging Completion, Circuit Malfunction |
| K.V.O Rate | 1 ml/h - 5 ml/h, default value 1 ml/h, can be programmed by user step 0.1 ml/h |
| Internal Battery | Lithium battery, 10.8V/2000mAh, more than 4 hours working back up |
| Power Consumption | 30 VA |
| Power | AC100V - 240V 50HZ/60HZ |
| Classification | Class II, Type CF, IPX4 |
| Dimension & Weight | 13×17.5×23 cm; 2 Kg |
| Optional Function | Ambulance DC12V |

| Parameter | SPA112 Syringe pump | SPA122 Syringe pump |
|---|---|---|
| Channel | Single | Double |
| Stackable | Yes | NO |
| Battery | ≥ 8 hours | ≥ 4 hours |
| Dimension (unit: cm) | 26X21.5X11 | 32X21.5X20 |
| Weight (unit: kg) | 2 | 3 |
| Syringe size (unit: ml) | 5, 10, 20, 30, 50/60 | |
| Infusion Mode | Rate Mode, Rate - Time, Rate - VTBI, Time - VTBI | |
| Safety | Double CPU to make sure higher standard infusion | |
| Accuracy | ±2% | |
| Time Preset | 00:01~99:59 (Hour:Min) | |
| Volume Range (unit: ml) | 0~9999.9 | |
| Flow Rate Step (unit: ml/h) | 0.1 when Rate < 100, 1 when Rate≥100 | |
| Flow Rate (unit: ml/h) | 5ml Syringe: 0.1 - 150 10ml Syringe: 0.1 - 300 20ml Syringe: 0.1 - 600 30ml Syringe: 0.1 - 900 50ml Syringe: 0.1 - 1200 |
|
| Purge/Bolus Rate/Top Flow Rate (unit: ml/h) | 5ml Syringe: 150 10ml Syringe: 300 20ml Syringe: 600 30ml Syringe: 900 50ml Syringe: 1200 |
|
| Audible & Visible Alarms | Auto Self - Test, Syringe Dislocation, Occlusion, Near End, Syringe Empty, VTBI Completion, Low Battery, Battery Exhausted, Motor Malfunction, Incorrect Syringe Specification, Circuit Malfunction, Main CPU Malfunction, Monitoring CPU Malfunction, Parameter Over Limitation, A.C Disconnection, A.C connection | |
| KVO (unit: ml/h) | 0.1 - 5.0 Adjustable | |
| Power Supply | AC 100 - 240V, 50hz - 60hz; DC 12V (Optional) | |
| Battery Type | Lithium battery, 11.1/2000mAh | |
| Classification | Class II, type CF, IPX4 | |