首页产品解决方案 / 输液泵

 

 

 

Model IPA112 Function Table

Function Details
I.V.Set Compatibility Open System, match all brands of qualified PVC, TPE I.V.sets with external diameter of 3.8mm - 4.2mm, Close system (Optional)
Range of Flow Rate 1 - 1200ml/h; 1 - 400d/min, step 1 drop
Infusion Mode Rate, Rate - Time, Rate - VTBI, Time - VTBI, Drop - VTBI, Drop - Time, Body Weight
Purge/Bolus Rate 1 - 1200ml/h, default 800ml/h, step by 1ml/h
Single Bolus Volume 1.0 - 10ml Adjustable, default 3 ml
Volume Range 1~9999ml
Accuracy ± 5%
Occlusion Pressure High 40 KPa; Middle 60 KPa; Low 100KPa (± 20KPa)
Alarms Door Open, Occlusion, VTBI Completion, Air in line, Error, Battery Exhaust, Pause overtime, Low Battery, Near Completion
K.V.O Rate 1ml/h - 5ml/h, default value 1ml/h, can be programmed by user step 0.1ml/h
Internal Battery Lithium battery, 10.8V/2200mAh, more than 4 hours working back up
Power Supply AC:100V - 240V 50HZ/60HZ; Ambulance DC: 12V(Optional)
Classification Class II, Type CF, IP34
Dimension & Weight 13 × 17.5 × 23 cm; 2.0Kg

 

 

 


 

 

产品信息表

Section Content
Product Name Infusion Pump IPD112
Slogan New Design New Experience
Feature 1 13 Infusion Modes
Rate, Rate-VTBI, Rate-time, Drop-VTBI, Time-VTBI, Body weight, Drug Library, Sequential, Loading dose, Trapezia, Intermittent, Mirco and Relay infusion function, TIVA (Optional)
Feature 2 High Accuracy
Flow Rate Step 0.01ml/h, Accuracy ±4%
Feature 3 DPS Function
DPS (Dynamic Pressure Detection) function, monitoring the pressure variations in real time.
Feature 4 UI Design
Friendly and clear for monitoring of infusion status with Rate, Time-left, VTBI, Accumulated volume, Pressure, Drug name etc.
Feature 5 User Customization
Drug library & Syringe brand can be modified according to clinical use.
Feature 6 Treatment Records
Can save up to 5 groups latest infusion solutions.
Feature 7 High Safety
Self-Test, Dual CPU, Up & Down Stream Occlusion Detection
Feature 8 Relay (Cascaded) infusion
Each units can be cascaded freely, no need other supplementary

Model IPD112 功能参数表

Function IPD112
LCD 3.5inch Touch Screen
Infusion Mode Rate, Rate-time, Drop-VTBI, Time-VTBI, Bodyweight, Drug Library, Sequential, Loading dose, Trapezia, Intermittent, Mirco and Relay infusion function, TIVA (optional)
Infusion Principle Fingertip Peristaltic Pump
I.V.Set Compatibility Open System, match all brands of qualified PVC, TPE I.V.sets with external diameter of 3.8mm - 4.3mm, Close system (Optional).
Range of Flow Rate 0.01 - 2200ml/h
Flow Rate Step 0.01ml/h
Purge/Bolus Rate 0.01 - 2200ml/h, default 800ml/h, step by 0.01ml/h
Single Bolus Volume 0.1 - 50ml Adjustable, default 3 ml
Time Preset 00:00:01~99:59:59 (H:M:S)
Volume Range 1~9999ml


 

产品信息表

Section Content
Product Name Infusion Pump IPA112 Plus
Slogan More Practical More Economical
Feature 1 Numeric key for user friendly and easy operation, operator can get how to use within 5mins
Feature 2 Open system to match different brands of I.V sets, Close system (Optional)
Feature 3 One-button night mode design for patient better rest at night
Feature 4 Multiple infusion modes of Rate, Rate-Time, Rate-VTBI, Time-VTBI, Body weight, Drug library to meet different clinical use
Feature 5 Drug library and Event Log 2000+ and can be customized
Feature 6 DPS (Dynamic Pressure Detection), more intelligent more safe

Model IPA112 Plus功能参数表

Function Model IPA112 Plus
LCD 3.5 inch LCD numeric key button
Infusion Mode Rate, Rate - Time, Rate - VTBI, Time - VTBI, Drop, Body Weight, Drug Library
Infusion Principle Fingertip Peristaltic Pump
I.V.Set Compatibility Open System, match all brands of qualified PVC, TPE I.V.sets with external diameter of 3.8mm - 4.2mm, Close system (Optional).
Range of Flow Rate 0.1 - 1500ml/h
Flow Rate Step 0.1 ml/h
Purge/Bolus Rate 1 - 1500ml/h, default 800ml/h, step by 1ml/h
Single Bolus Volume 1.0 - 10ml Adjustable, default 3 ml
Time Preset 00:01~99:59 (Hour: Min)
Volume Range 1~9999ml
Drop Range 1 - 500d/min, step by 1 drop
Accuracy ±5%
Air in line Detection Ultrasonic Wave
Alarms and Prompts Air in line, Occlusion, Door Open, VTBI Completion, Near Completion, Low Battery, Battery Exhausted, Low Temperature, Motor Malfunction, A.C Disconnection, A.C connection, I.V set dislocation, MPU Error, Battery Charging, Battery Charging Completion, Circuit Malfunction
KVO 1ml/h - 5ml/h, default value 1ml/h, can be programmed by user step 0.1ml/h
Occlusion Pressure Low 40KPa ± 20KPa
Middle 60KPa ± 20KPa
High 100KPa ± 20KPa
Internal Battery Lithium battery,10.8V/2200mAh, more than 4 hours working back up
Power Supply 30VA; AC100V - 240V 50HZ/60HZ
Classification Class II, type CF, IP34
Dimension & Weight 13*17.5*23cm, 2KG
Optional Function Ambulance DC12V, WIFI, Language customization